EchoPod®
UG01 & UG03

EchoPod-UG01-UG03-Vertical-360

Condensation is the most common variable in liquid level applications. Condensation attenuates the acoustic signal of ultrasonic sensors with horizontal transducers, weakening their signal strength and signal to noise ratio by up to 50%, and substantially reducing their measurement reliability. At the core of Reflective Technology™ is a simple fact. Unlike flat horizontal surfaces, significant water droplets cannot adhere to smooth vertical surfaces. By orienting the transducer vertically, condensation runs off the transducer and does not affect sensor performance. The unimpeded transmit and receive signals are redirected to and from the liquid off a 45º reflector, delivering reliable level measurement. Thanks gravity.

Reflective Transducer
Reflective-Transducer-UG01-03
Patent Pending
Reliable Measurement
Horizontal Transducer
Horizontal-Transducer-UG01-03
Signal Attenuation
Specification

Range: UG01: 1.5” to 4.9’ (3.8cm to 1.5m), UG03: 4” to 9.8’ (10.1cm to 3m)

Accuracy: UG01: 0.125” (3mm), UG03: ± 0.2% of range

Resolution: UG01: 0.019” (0.5mm), UG03: 0.039” (1mm)

Dead band: UG01: 1.5” (3.8cm), UG03: 4” (10.1cm)

Beam width: 2” (5cm)

Configuration: WebCal® PC Windows® USB® 2.0

Memory: Non-volatile

Supply voltage: 14-28 VDC

Max. consumption: 0.5W

Loop resistance: 500 ohms @ 24 VDC

Signal output: 4-20mA, two-wire

Signal invert: 4-20mA or 20-4mA

Signal fail-safe: 4mA, 20mA, 21mA, 22mA, hold last

Contact type: (4) SPST relays

Contact rating: 60 VA, 1A maximum

Contact fail-safe: Open, closed, hold last

Contact hysteresis: Selectable

Process temp.: F: -40º to 176º, C: -40º to 80º

Temp. comp.: Automatic

Ambient temp.: F: -31º to 140º, C: -35º to 60º

Pressure: 30 psi (2 bar)

Enclosure rating: Type 6P (IP68)

Encl. material: Polypropylene

Transducer type: Reflective

Transducer mat.: Polyvinylidene fluoride

Cable jacket mat.: Polyurethane

Cable type: 9-conductor, shielded

Cable length: 48” (1.2m)

Process mount: -0001: 2” NPT, -0011: 2” G

Mount gasket: -0001: N/a, -0011: Viton®

Classification: General purpose

Approval: UL 61010-1

Compliance: CE, RoHS

EchoPod®
UG06 & UG012

ug06-ug12

Condensation is the most common variable in liquid level applications. Condensation attenuates the acoustic signal of ultrasonic sensors with horizontal transducers, weakening their signal strength and signal to noise ratio by up to 50%, and substantially reducing their measurement reliability. At the core of Reflective Technology™ is a simple fact. Unlike flat horizontal surfaces, significant water droplets cannot adhere to smooth vertical surfaces. By orienting the transducer vertically, condensation runs off the transducer and does not affect sensor performance. The unimpeded transmit and receive signals are redirected to and from the liquid off a 45º reflector, delivering reliable level measurement. Thanks gravity.

Reflective Transducer
Reflective-Transducer-UG01-03
Patent Pending
Reliable Measurement
Horizontal Transducer
Horizontal-Transducer-UG01-03
Signal Attenuation
Specification

Range: UG06: 8” to 19.6’ (20cm to 6m), UG12: 18” to 39.3’ (45.7cm to 12m)

Accuracy: ± 0.2% of range

Resolution: UG06: 0.079” (2mm), UG12: 0.196” (5mm)

Dead band: UG06: 8” (20.3cm), UG12: 18” (45.7cm)

Beam width: UG06: 3” (7.6cm), UG12: 6” (15.2cm)

Configuration: Push button or WebCal® PC Windows® USB® 2.0

Memory: Non-volatile

Display type: LCD, 6-digit

Display units: Inch, cm or percent

Supply voltage: 14-28 VDC

Max. consumption: 0.5W

Loop resistance: 500 ohms @ 24 VDC

Signal output: 4-20mA, two-wire

Signal invert: 4-20mA or 20-4mA

Signal fail-safe: 4mA, 20mA, 21mA, 22mA, hold last

Process temp.: F: -40º to 176º, C: -40º to 80º

Temp. comp.: Automatic

Ambient temp.: F: -31º to 140º, C: -35º to 60º

Pressure: 30 psi (2 bar)

Enclosure rating: IP68

Encl. material: Polypropylene

Encl. cap material: Clear polycarbonate

Enclosure vent: Water tight membrane

Conduit entrance: 1/2” NPT

Transducer type: Reflective

Transducer mat.: Polyvinylidene fluoride

Process mount: 06-0001: 2” NPT, 06-0011: 2” G, 12-0001: 3″ NPT, 12-0011: 3″ G

Mount gasket: -0001: N/a, -0011: Viton®

Classification: General purpose

Approval: UL 61010-1

Compliance: CE, RoHS

EchoPod®
DL14

ultrasonic-600x600

The general purpose ultrasonic multi-function level transmitter provides continuous level measurement up to 49.2” (1.25m) with a 4-20 mA analog signal output and four relays, and is configured via our WebCal software. Each relay can be configured for alarm, automatic fill or empty functions. Select this sensor for small tanks with non-foaming, non-condensing or highly vaporous media such as chemicals, water, wastewater and oil. Typical applications include chemical feed, skid or machine, IBC or drum and small waste sumps.

Specification

Range: 49.2” (1.25m)

Accuracy: 0.125” (3mm)

Resolution: 0.019” (0.5mm)

Dead band: 2” (5cm)

Beam width: 2” (5cm)

Configuration: WebCal™ PC, Windows®, USB 2.0

Memory: Non-volatile

Supply voltage: 24 VDC (loop)

Consumption: 0.5W

Loop resist.: 400Ω max

Signal output: 4-20 mA, two-wire

Signal invert: 4-20 mA or 20-4mA

Signal fail-safe: 4 mA, 20 mA, 21 mA, 22 mA or hold last

Contact type: (4) SPST relays, 1A

Contact fail-safe: Power loss: Hold last, Power on: Open, close or hold last

Hysteresis: Selectable

Process temp.: F: 20° to 140°, C: -7° to 60°

Temp. comp.: Automatic

Ambient temp.: F: -31° to 140°, C: -35° to 60°

Pressure: MWP = 30 PSI (2 bar)

Enclosure rating: Type 6P, encapsulated, corrosion resistant & submersible

Encl. material: Polycarbonate

Strain relief mat.: Santoprene

Trans. material: PVDF

Cable jacket mat.: Polyurethane

Cable type: 9-conductor, shielded

Cable length: 48” (1.2m)

Process mount: 1” NPT (1” G)

Mount gasket: Viton®

Classification: General purpose

Compliance: CE, RoHS

Approvals: cFMus

Download Datasheet

EchoPod® DL14 Multi-Function Ultrasonic Liquid Level Sensor

 

EchoPod®
DL10

zoom-ultrasonic-level-transmitter-echopod-dl10-98491_orig

The general purpose ultrasonic level transmitter provides continuous level measurement up to 49.2” (1.25m) with a 4-20 mA signal output, and is configured via our WebCal software. Select this sensor for small tanks with non-foaming, non-condensing or highly vaporous media such as chemical, water, wastewater and oil. Typical applications include chemical feed, skid or machine, IBC or drum and small waste sumps.

Specification

Range: 49.2” (1.25m)

Accuracy: 0.125” (3mm)

Resolution: 0.019” (0.5mm)

Dead band: 2” (5cm)

Beam width: 2” (5cm)

Configuration: WebCal™ PC, Windows®, USB 2.0

Memory: Non-volatile

Supply voltage: 24 VDC (loop)

Consumption: 0.5W

Loop resist.: 400Ω max

Signal output: 4-20 mA, two-wire

Signal invert: 4-20 mA or 20-4mA

Loop fail-safe: 4 mA, 20 mA, 21 mA, 22 mA or hold last

Process temp.: F: 20° to 140°, C: -7° to 60°

Temp. comp.: Automatic

Ambient temp.: F: -31° to 140°, C: -35° to 60°

Pressure: MWP = 30 PSI (2 bar)

Enclosure rating: Type 6P, encapsulated, corrosion resistant & submersible

Encl. material: Polycarbonate

Strain relief mat.: Santoprene

Trans. material: PVDF

Cable jacket mat.: Polyurethane

Cable type: 4-conductor, shielded

Cable length: 48” (1.2m)

Process mount: 1” NPT (1” G)

Mount gasket: Viton®

Classification: General purpose

Compliance: CE, RoHS

Approvals: cFMus